New TopShim run:

version = topshim 2.0.0
date = 2025-06-08
time = 09:55:59.877
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0631
solvent = CD3CN
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 1.93 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

initial B0 stdDev = 0.56 Hz
sample size = 3.00 cm, position = -0.04 cm
final B0 stdDev = 0.40 Hz > improvement = 1.4
envelope width = 0.54 Hz

Performing tune after shimming
tune lock level increase = 1.1

shim changes: 
Z	-25
Z2	+3
Z3	+14
Z4	+3
Z5	-34
X	-16
Y	-4

duration = 2 min 33 sec
completed successfully
finished Sun Jun  8 09:58:30 2025


========================================================================